AirChek Essential Air Sampling Pump

Easy to use 5 to 5000ml/min Air Sampling Pump

SKU
ACESSENTIAL

AirChek Essential Touch PadThe SKC AirChek Essential Sample Pump is the latest SKC Sample Pump in the AirChek range. The SKC AirChek Essential is ideal for users who prefer the basic operation principle of SKC’s Basic Pumps such as the AirChek 52, the AirLite Series and even the XR5000 Sampling Pump but don’t have a need for datalogging and the extra features offered in SKC’s premier pump, the AirChek TOUCH.

Like the AirChek TOUCH Sample Pump, the intrinsically safe AirChek Essential features a new packaging, increased durability, higher flow range (5-5000 ml/min) and also offers longer run times of up to 40+ hours at 2000 ml/min. With the ability to correct flow for changes in temperature and atmospheric pressure, the AirChek Essential also features tough no-slip anti-static case, bright running and status LEDs and an intuitive touch keypad operation.

For more information about the SKC AirChek Essential Sampling Pump, contact your local Air-Met Scientific office today.

SKC AirChek Essential Air Sampling Pump Features

  • Flows from 5 to 5000 ml/min 
    • 5 to 500 ml/min with low flow holder accessory 
    • Multiple-tube sampling with low flow accessories 
  • High back pressure compensation 
    • Up to 50 inches water at 2 L/min
  • Li-Ion battery — over 40-hour run times 
  • Corrects flow for changes in temperature and atmospheric pressure 
  • Bright running and status LEDs
  • Displays elapsed time
  • Flow fault and auto-restart 
  • Tough no-slip anti-static case
  • Ultra-quiet
  • Small and lightweight
  • Convenient charging cradle
    • Chain up to 5 cradles
  • Screen protection and security 
    • Protective cover 
    • Screen lock-out feature

SKC AirChek Essential Air Sampling Pump Specifications

Flow Range Constant flow from 1000 to 5000 ml/min (5 to 500 ml/min requires low flow holder)
Compensation Range (Back Pressure Capability) 
  • 5000 ml/min at 20 inches water back pressure 
  • 4000 ml/min at 30 inches water back pressure
  • 3000 ml/min at 40 inches water back pressure
  • 2000 ml/min at 50 inches water back pressure
  • 1000 ml/min at 60 inches water back pressure
Power Removable rechargeable lithium-ion (Li-Ion) battery, 7.4 V, 2.6 Ah, 19.2 Wh or AC using cradle
Run Time 40+ hours at 2000 ml/min,* 15+ hours at 5000 ml/ min,* Indefinite run from charging cradle
Accuracy
  • Flow control: ± 5% of set-point after calibration to desired flow 
  • Atmospheric pressure: ± 0.3 in Hg
  • Temperature: ± 1°C
Display/Parameters B&W display, elapsed sampling time
User Interface Capacitive touch keypad with locking option
Sound Level  Average 51.7 dB at 3-ft (1-m) distance using a 37-mm, 0.8-µm MCE filter cassette
Tubing Requires ¼-inch ID tubing
Dimensions/Weight 4.1 x 3.7 x 2.8 in (10.4 x 9.4 x 7.1 cm), 19.4 oz (550 gm)
Certifications
  • Intrinsic safety (SKC Cat. No. 220-3000 operated with SKC Battery Pack Cat. No. P75718) 
  • Class I, Groups A, B, C, D; Class II, Groups E, F, G; Class III, T4; Class I, Zone 0
  • Aexia IIC T4 Ga; Exia IIC T4 Ga; -20 C = Ta = 45 C Ex II 1G; IECEx UL 19.0100 
  • DEMKO 19ATEX 2288 
  • CE 0539 
  • RoHS compliant 
  • Designed to meet ISO 13137:2013